Skip to main navigation Skip to search Skip to main content

Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry

  • G. Mazza
  • , R. Wheadon
  • , F. Lenta
  • , C. Buti
  • , M. Fabi
  • , C. Grimani
  • , S. Pallotta
  • , F. Sabbatini
  • , C. Talamonti
  • , M. Villani
  • , S. Aziz
  • , L. Calcagnile
  • , A. P. Caricato
  • , M. Martino
  • , G. Maruccio
  • , A. G. Monteduro
  • , G. Quarta
  • , S. Rizzato
  • , R. Catalano
  • , G. A.P. Cirrone
  • G. Cuttone, M. C. Guarrera, G. Petringa, L. Frontini, V. Liberali, A. Stabile, M. Vasquez, T. Croci, M. Ionica, K. Kanxheri, M. Menichelli, A. Morozzi, F. Moscatelli, D. Passeri, M. Pedio, F. Peverini, P. Placidi, L. Servoli, L. Tosti, N. Zema, D. Caputo, G. De Cesare, N. Lovecchio, A. Nascetti, D. Calvo, R. Cirio, P. De Remigis, S. Dunand, J. E. Thomet, M. Large, M. Petasecca, A. Bashiri, L. Piccolo, N. Wyrsch
  • INFN Sezione di Torino
  • Politecnico di Torino
  • Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
  • University of Florence
  • University of Urbino Carlo Bo
  • Sezione di Lecce
  • University of Salento
  • INFN-LNS
  • Sezione INFN di Milano
  • University of Milano
  • INFN Sezione di Perugia
  • University of Perugia
  • International School for Advanced Studies (SISSA/ISAS)
  • Consiglio Nazionale delle Ricerche
  • Sezione di Roma
  • University of Rome “Tor Vergata”
  • ENAC-IIC-GEL
  • University of Wollongong
  • Najran University

Research output: Contribution to journalArticlepeer-review

Abstract

The Cleopatra ASIC is a 12-channel prototype ASIC for the readout of hydrogenated amorphous silicon sensors used for real-time dosimetry in radiation diagnostic and radiation therapy. The architecture is based on a current to frequency conversion based on the recycling integrator principle in order to cover a dynamic range of four orders of magnitude with high linearity. Three different input amplifier configurations have been implemented in order to check the trade-off between detector capacitance and maximum output frequency. Cleopatra has been designed in CMOS 28 nm technology and succesfully tested in laboratory.

Original languageEnglish
Article numberC01034
JournalJournal of Instrumentation
Volume20
Issue number1
DOIs
Publication statusPublished - 1 Jan 2025
Externally publishedYes

Keywords

  • Analogue electronic circuits
  • Digital electronic circuits
  • Dosimetry concepts and apparatus
  • Front-end electronics for detector readout

Fingerprint

Dive into the research topics of 'Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry'. Together they form a unique fingerprint.

Cite this