Abstract
The Cleopatra ASIC is a 12-channel prototype ASIC for the readout of hydrogenated amorphous silicon sensors used for real-time dosimetry in radiation diagnostic and radiation therapy. The architecture is based on a current to frequency conversion based on the recycling integrator principle in order to cover a dynamic range of four orders of magnitude with high linearity. Three different input amplifier configurations have been implemented in order to check the trade-off between detector capacitance and maximum output frequency. Cleopatra has been designed in CMOS 28 nm technology and succesfully tested in laboratory.
| Original language | English |
|---|---|
| Article number | C01034 |
| Journal | Journal of Instrumentation |
| Volume | 20 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 1 Jan 2025 |
| Externally published | Yes |
Keywords
- Analogue electronic circuits
- Digital electronic circuits
- Dosimetry concepts and apparatus
- Front-end electronics for detector readout
Fingerprint
Dive into the research topics of 'Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry'. Together they form a unique fingerprint.Cite this