Extended eigenvalue calibration method for overdetermined Mueller matrix polarimeters

Research output: Contribution to journalLetterpeer-review

Abstract

The eigenvalue calibration method is a versatile approach that can be applied to any type of the Mueller matrix polarimetic setup because a precise knowledge of the optical response of the setup components is not required. The method has usually been employed in its original form to calibrate non-overdetermined polarimeters dealing with intensity data arranged in 4×4 matrices, but it can also be applied to calibrate overdetermined polarimeters with intensity data matrices of higher dimension. The main drawback with the original formulation of the method is its sensitivity to noise in the input data, especially if applied as it is to overdetermined intensity matrices. In the present work, we present a rigorous extension of the conventional eigenvalue calibration method to treat overdetermined data.We experimentally show that the proposed method does not enhance noise propagation, and therefore it does not degrade the quality of Mueller matrices measured with overdetermined polarimeters.

Original languageEnglish
Pages (from-to)1165-1168
Number of pages4
JournalOptics Letters
Volume49
Issue number5
DOIs
Publication statusPublished - 1 Mar 2024

Fingerprint

Dive into the research topics of 'Extended eigenvalue calibration method for overdetermined Mueller matrix polarimeters'. Together they form a unique fingerprint.