Force microscopy cantilevers locally heated in a fluid: Temperature fields and effects on the dynamics

Basile Pottier, Ludovic Bellon

Research output: Contribution to journalArticlepeer-review

Abstract

Atomic force microscopy cantilevers are often, intentionally or not, heated at their extremity. We describe a model to compute the resulting temperature field in the cantilever and in the surrounding fluid on a wide temperature range. In air and for common geometries, the heat fluxes in the cantilever and to the environment are of comparable magnitude. We then infer how the fluid-structure interaction is modified due to heating and predict the induced changes in the dynamics of the system. In particular, we describe how the resonance frequencies of the cantilever shift with a temperature increase due to two competing processes: softening of the cantilever and decrease of the fluid inertial effects. Our models are illustrated by experiments on a set of cantilevers spanning the relevant geometries to explore the relative importance of both effects.

Original languageEnglish
Article numbere0060911
JournalJournal of Applied Physics
Volume130
Issue number12
DOIs
Publication statusPublished - 28 Sept 2021
Externally publishedYes

Fingerprint

Dive into the research topics of 'Force microscopy cantilevers locally heated in a fluid: Temperature fields and effects on the dynamics'. Together they form a unique fingerprint.

Cite this