Skip to main navigation Skip to search Skip to main content

HGCROC3: The front-end readout ASIC for the CMS High Granularity Calorimeter

  • F. Bouyjou
  • , G. Bombardi
  • , F. Dulucq
  • , A. El Berni
  • , S. Extier
  • , M. Firlej
  • , T. Fiutowski
  • , F. Guilloux
  • , M. Idzik
  • , C. De La Taille
  • , A. Marchioro
  • , A. Molenda
  • , J. Moron
  • , K. Swientek
  • , D. Thienpont
  • , T. Vergine
  • Universite Paris-Saclay
  • European Organization for Nuclear Research
  • Omega (Centre de Microélectronique)-École Polytechnique
  • Agh University of Science and Technology Faculty of Computer Science

Research output: Contribution to journalArticlepeer-review

23 Citations (Scopus)

Abstract

For the CMS High Granularity Calorimeter (CE), the final version of the 72-channel front-end ASIC (HGCROC3) was submitted in December 2020. HGCROC3 includes low-noise/high-gain preamplifiers/shapers and a 10-bit 40 MHz successive approximation ADC (SAR-ADC) that provide the charge measurement over the linear range of the preamplifier. In the saturation range, a discriminator and a time-to-digital converter (TDC) provide the charge information from the time over threshold (ToT; 200 ns dynamic range, 50 ps binning). A fast discriminator and another TDC provide timing information to 25 ps accuracy. The chip embeds all necessary ancillary services: bandgap circuit, PLL, threshold DACs. We present the experimental results on the latest and final version (HGCROC3) received in April 2021.

Original languageEnglish
Article numberC03015
JournalJournal of Instrumentation
Volume17
Issue number3
DOIs
Publication statusPublished - 1 Mar 2022
Externally publishedYes

Keywords

  • Front-end electronics for detector readout
  • Radiation-hard electronics

Fingerprint

Dive into the research topics of 'HGCROC3: The front-end readout ASIC for the CMS High Granularity Calorimeter'. Together they form a unique fingerprint.

Cite this