Skip to main navigation Skip to search Skip to main content

Integrated setup for time and spatially resolved micro-photoluminescence

  • M. Perlangeli
  • , G. M. Pierantozzi
  • , A. Fondacaro
  • , P. Sander
  • , E. Dobovicnik
  • , F. Parmigiani
  • , G. Rossi
  • , G. Panaccione
  • , F. Cilento
  • , R. Cucini
  • International School for Advanced Studies (SISSA/ISAS)
  • University of Trieste
  • Université de Strasbourg
  • Nova Gorica Univ., Slovenia and Sincrotrone ELETTRA
  • International Faculty
  • Germany; University of Cologne
  • Dipartimento di Fisica
  • University of Milano

Research output: Contribution to journal › Article › peer-review

Abstract

We report on a setup for photoluminescence (PL) spectroscopy, allowing to perform time resolved, steady-state, and spatially resolved PL and PL imaging. Time domain information (with ≈50 ps resolution) is achieved via time-correlated single photon counting (TCSPC), while the spectral information (in the ≈1–4 eV NIR-UV energy range) is accessed through a common-path interferometer based on birefringence, with the PL spectrum being retrieved via a Fourier transform approach. The micro-PL approach allows us to measure the PL signal of micron-sized samples, such as exfoliated 2D semiconductors. Moreover, the high efficiency of signal collection and analysis allows us to resolve weak PL signals from indirect-bandgap semiconductors. We measure the time-, frequency-, and spatially resolved PL on exfoliated flakes of WSe2, MoS2, and WS2 as examples to display the potential of our new apparatus.

Original languageEnglish
Article number023004
JournalReview of Scientific Instruments
Volume97
Issue number2
DOIs
Publication statusPublished - 1 Feb 2026
Externally publishedYes

Fingerprint

Dive into the research topics of 'Integrated setup for time and spatially resolved micro-photoluminescence'. Together they form a unique fingerprint.

Cite this