Narrow Aperture Detection Electrodes ICR Cell with Quadrupolar Ion Detection for FT-ICR MS at the Cyclotron Frequency

  • Konstantin O. Nagornov
  • , Anton N. Kozhinov
  • , Edith Nicol
  • , Oleg Yu Tsybin
  • , David Touboul
  • , Alain Brunelle
  • , Yury O. Tsybin

Research output: Contribution to journalArticlepeer-review

Abstract

Ion signal detection at the true (unperturbed) cyclotron frequency instead of the conventional reduced cyclotron frequency has remained a formidable challenge since the inception of Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS). Recently, routine FT-ICR MS at the true cyclotron frequency has become a reality with the implementation of ICR cells with narrow aperture detection electrodes (NADEL). Here, we describe the development and implementation of the next generation of these cells, namely, a 2xNADEL ICR cell, which comprises four flat detect and four ∼45° cylindrical excite electrodes, enabling independent ion excitation and quadrupolar ion detection. The performance of the 2xNADEL ICR cell was evaluated on two commercial FT-ICR MS platforms, 10 T LTQ FT from Thermo Scientific and 9.4 T SolariX XR from Bruker Daltonics. The cells provided accurate mass measurements in the analyses of singly and multiply charged peptides (root-mean-square, RMS, mass error Δm/m of 90 ppb), proteins (Δm/m = 200 ppb), and petroleum fractions (Δm/m < 200 ppb). Due to the reduced influence of measured frequency on the space charge and external (trapping) electric fields, the 2xNADEL ICR cells exhibited stable performance in a wide range of trapping potentials (1-20 V). Similarly, in a 13 h rat brain MALDI imaging experiment, the RMS mass error did not exceed 600 ppb even for low signal-to-noise ratio analyte peaks. Notably, the same set of calibration constants was applicable to Fourier spectra in all pixels, reducing the need for recalibration at the individual pixel level. Overall, these results support further experimental development and fundamentals investigation of this promising technology.

Original languageEnglish
Pages (from-to)2258-2269
Number of pages12
JournalJournal of the American Society for Mass Spectrometry
Volume31
Issue number11
DOIs
Publication statusPublished - 4 Nov 2020
Externally publishedYes

Keywords

  • FT
  • FTMS
  • Fourier transform
  • Fourier transform mass spectrometry
  • ICR
  • NADEL
  • cyclotron frequency
  • ion cyclotron resonance
  • magnetron frequency
  • narrow aperture detection electrode
  • quadrupolar ion detection
  • reduced cyclotron frequency

Fingerprint

Dive into the research topics of 'Narrow Aperture Detection Electrodes ICR Cell with Quadrupolar Ion Detection for FT-ICR MS at the Cyclotron Frequency'. Together they form a unique fingerprint.

Cite this