Skip to main navigation Skip to search Skip to main content

New compact femtosecond laser source for penetrating keratoplasty at 1.65 μm

  • C. Crotti
  • , F. Deloison
  • , D. A. Peyrot
  • , F. Aptel
  • , F. Alahyane
  • , L. Kowalczuk
  • , T. Marciano
  • , K. Plamann
  • , M. Savoldelli
  • , J. M. Legeais
  • Laboratory d'Optique Appliquée, ENSTA, CNRS-École Polytechnique
  • ALPhANOV
  • Centre national de la recherche scientifique
  • Hôpital Édouard Herriot
  • Clinical Epidemiology Centre

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

Femtosecond laser surgery in the volume of corneal tissue is typically performed wavelengths of about 1 μm, which gives excellent results on transparent corneas. However, the outcome is much worse in the case of oedematous or pathological corneas as the laser beam propagation is disturbed by optical scattering. Our studies suggest that this phenomenon can be greatly reduced by using a better suited laser wavelength. Best results are obtained at 1.65 μm. Currently, no compact femtosecond laser at this wavelength is commercially available. We have developed a new simple, compact and stable laser source consisting of a non linear crystal pumped by a compact commercial solid-state laser emitting at 1.03 μm in a configuration of an Optical Parametric Generation (OPG). The output wavelength of this system can be tuned in the spectral range of 1.45 - 1.8 μm. A series of ex vivo penetrating incisions using energies of the order of a few microjoules on corneal tissues have been performed while varying the wavelengths from 1.45 μm to 1.7 μm. The results have been compared to experiments performed at 0.8 μm and 1 μm. The use of longer infrared wavelengths around 1.65 μm for femtosecond laser keratoplasty significantly improves the quality and the penetration depth of incision in case of pathological tissues, without inducing any additional side effects.

Original languageEnglish
Title of host publicationEuropean Conference on Biomedical Optics, ECBO 2011
Publication statusPublished - 1 Dec 2011
EventEuropean Conference on Biomedical Optics, ECBO 2011 - Munich, Germany
Duration: 22 May 201126 May 2011

Publication series

NameOptics InfoBase Conference Papers
ISSN (Electronic)2162-2701

Conference

ConferenceEuropean Conference on Biomedical Optics, ECBO 2011
Country/TerritoryGermany
CityMunich
Period22/05/1126/05/11

Keywords

  • Absorption
  • Corneal surgery
  • Scattering
  • Suited wavelength
  • Transparency

Fingerprint

Dive into the research topics of 'New compact femtosecond laser source for penetrating keratoplasty at 1.65 μm'. Together they form a unique fingerprint.

Cite this