Skip to main navigation Skip to search Skip to main content

Sub-kHz-level relative stabilization of an intracavity doubled continuous wave optical parametric oscillator using Pound-Drever-Hall scheme

  • Laboratoire Aimé Cotton

Research output: Contribution to journalArticlepeer-review

Abstract

We report the relative frequency stabilization of an intracavity frequency doubled singly resonant optical parametric oscillator on a Fabry-Perot ́etalon. The red/orange radiation produced by the frequency doubling of the intracavity resonant idler is stabilized using the Pound-Drever-Hall locking technique. The relative frequency noise of this orange light, when integrated from 1 Hz to 50 kHz, corresponds to a standard deviation of 700 Hz. The frequency noise of the pump laser is shown experimentally to be transferred to the non resonant signal beam.

Original languageEnglish
Pages (from-to)18049-18057
Number of pages9
JournalOptics Express
Volume19
Issue number19
DOIs
Publication statusPublished - 12 Sept 2011
Externally publishedYes

Fingerprint

Dive into the research topics of 'Sub-kHz-level relative stabilization of an intracavity doubled continuous wave optical parametric oscillator using Pound-Drever-Hall scheme'. Together they form a unique fingerprint.

Cite this