Résumé
We present the use of a low background counting facility, equipped with a p-type 80% relative efficiency HPGe detector, protected by active and passive shielding, and large enough to count a 10 in photo-multiplier tube (PMT). A GEANT4 Monte-Carlo of this detector was developed and tuned to 3% accuracy. We report the U, Th, and K content in three different types of PMTs used in current neutrino experiments, with accuracies of -10ppb for U and Th and of -15ppm for K.
| langue originale | Anglais |
|---|---|
| Pages (de - à) | 1033-1038 |
| Nombre de pages | 6 |
| journal | Applied Radiation and Isotopes |
| Volume | 69 |
| Numéro de publication | 7 |
| Les DOIs | |
| état | Publié - 1 juil. 2011 |
| Modification externe | Oui |
Empreinte digitale
Examiner les sujets de recherche de « A large HPGe detector for the non-destructive radioassay of an ultra-low-background counting facility ». Ensemble, ils forment une empreinte digitale unique.Contient cette citation
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver