Passer à la navigation principale Passer à la recherche Passer au contenu principal

A new robust method for six-port reflectometer calibration

Résultats de recherche: Contribution à un journalArticleRevue par des pairs

Résumé

A new robust method for finding the parameters of Engen's six-port-to-four-port reduction algorithm for six-port reflectometer calibration has been developed. Like other previously published methods, it uses a minimum of five loads with an unknown but constant absolute value of the reflection coefficient and unknown but well-distributed phases. However, the quality of the parameter estimates is improved, especially in noisy environments, by efficiently eliminating cases in which these earlier methods may become ill-conditioned. The new method has been used successfully to calibrate a newly developed six-port reflectometer in GaAs MMIC technology working between 1.3 GHz and 3.0 GHz.

langue originaleAnglais
Pages (de - à)927-931
Nombre de pages5
journalIEEE Transactions on Instrumentation and Measurement
Volume48
Numéro de publication5
Les DOIs
étatPublié - 1 oct. 1999

Empreinte digitale

Examiner les sujets de recherche de « A new robust method for six-port reflectometer calibration ». Ensemble, ils forment une empreinte digitale unique.

Contient cette citation