Résumé
We demonstrate selective two-photon excited fluorescence microscopy with shaped pulses produced with a simple yet efficient scheme based on dispersive optical components. The pulse train from a broadband oscillator is split into two subtrains that are sent through different amounts of glass. Beam recombination results in pulse-shape switching at a rate of 150 MHz. Time-resolved photon counting detection then provides two simultaneous images resulting from selective two-photon excitation, as demonstrated in a live embryo. Although less versatile than programmable pulse-shaping devices, this novel arrangement significantly improves the performance of selective microscopy using broadband shaped pulses while simplifying the experimental setup.
| langue originale | Anglais |
|---|---|
| Pages (de - à) | 3444-3446 |
| Nombre de pages | 3 |
| journal | Optics Letters |
| Volume | 35 |
| Numéro de publication | 20 |
| Les DOIs | |
| état | Publié - 15 oct. 2010 |