First tests in the application of silicon photomultiplier arrays to dose monitoring in hadron therapy

  • CSIC-UV - Instituto de Física Corpuscular (IFIC)
  • Laboratoire de l'Accélérateur Linéaire

Résultats de recherche: Contribution à un journalArticleRevue par des pairs

13 Citations (Scopus)

Résumé

A detector head composed of a continuous LaBr3 crystal coupled to a silicon photomultiplier array has been mounted and tested, for its use in a Compton telescope for dose monitoring in hadron therapy. The LaBr3 crystal has 16 mm×18 mm×5 mm size, and it is surrounded with reflecting material in five faces. The SiPM array has 16 (4 ×4) elements of 3 mm×3 mm size. The SPIROC1 ASIC has been employed as readout electronics. The detector shows a linear behavior up to 1275 keV. The energy resolution obtained at 511 keV is 7% FWHM, and it varies as one over the square root of the energy up to the energies tested. The variations among the detector channels are within 12%. A preliminary measurement of the timing resolution gives 7 ns FWHM. The spatial resolution obtained with the center of gravity method is 1.2 mm FWHM. The tests performed confirm the correct functioning of the detector.

langue originaleAnglais
Pages (de - à)S96-S99
journalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume648
Numéro de publicationSUPPL. 1
Les DOIs
étatPublié - 21 août 2011
Modification externeOui

Empreinte digitale

Examiner les sujets de recherche de « First tests in the application of silicon photomultiplier arrays to dose monitoring in hadron therapy ». Ensemble, ils forment une empreinte digitale unique.

Contient cette citation