Minimizing Disturbance in Ion Beam Profiling with PEPITES Monitor

  • A. Esper
  • , L. Bernardi
  • , M. Donetti
  • , R. Duhamel
  • , F. Gastaldi
  • , O. Gevin
  • , R. Guillaumat
  • , M. Pullia
  • , C. Thiebaux
  • , M. Verderi
  • , C. Viviani
  • Ip Paris
  • CNAO National Center for Oncological Hadrontherapy
  • Université Paris-Saclay

Résultats de recherche: Contribution à un journalArticle de conférenceRevue par des pairs

Résumé

The PEPITES profiler is a novel beam diagnostic tool that utilizes secondary electron emission (SEE) designed for precise and low-disturbance profiling of charged particle beams. Reducing the monitor WET (Water Equivalent Thickness) allows to minimize the disturbance on the beam, extending the PEPITES usage to lower energies and to farther monitor-target distance applications. This study compares the performances of two detector configurations: the original In-axis anode design, featuring a parallel electron collection electric field, and the upgraded Off-axis anode design, which offers a reduced WET, but with a non-parallel field. Results show that the Off-axis design provides accurate measurements, unaffected by the modified field. These improvements confirm the PEPITES profiler as a promising tool for high-precision beam diagnostics for hadron-therapy.

langue originaleAnglais
Numéro d'article012043
journalJournal of Physics: Conference Series
Volume3094
Numéro de publication1
Les DOIs
étatPublié - 1 janv. 2025
Evénement16th International Particle Accelerator Conference, IPAC 2025 - Taipei, Taiwan
Durée: 1 juin 20256 juin 2025

Empreinte digitale

Examiner les sujets de recherche de « Minimizing Disturbance in Ion Beam Profiling with PEPITES Monitor ». Ensemble, ils forment une empreinte digitale unique.

Contient cette citation