The design of a six-port reflectometer with frequency-independent calibration procedure

  • France

Résultats de recherche: Contribution à un journalArticle de conférenceRevue par des pairs

3 Citations (Scopus)

Résumé

The accuracy in the measurement of an unknown reflection coefficient using a six-port automatic network analyzer (ANA) depends in part on the stability of the calibration constants of the system. A six-port reflectometer whose configuration minimizes the variations of these parameters with frequency is described. The advantages of a six-port configuration include a weakly and slowly frequency-dependent calibration process. Consequently, uncertainties due to the variations of the six-port parameters with RF-source stability are reduced. There is no need for a synthesized source, and the number of calibration points can be reduced while the symmetrical distribution of the qi points can be kept over the whole frequency range.

langue originaleAnglais
Pages (de - à)398-399
Nombre de pages2
journalCPEM Digest (Conference on Precision Electromagnetic Measurements)
étatPublié - 1 juin 1990
Modification externeOui
Evénement1990 Conference on Precision Electromagnetic Measurements - Ottawa, Ont, Can
Durée: 11 juin 199014 juin 1990